Photovac Voyager
Portable Gas Chromatograph
Featured Rental
An extremely powerful portable gas chromatograph, the Voyager features a 3 column design and optional dual PID/ECD detectors for advanced selective gas analysis. The Voyager GC is fully self contained with an internal rechargeable battery and onboard carrier gas cylinder. Voyager rentals are available with standard applications for Environmental, ABS rubber, or petrochemical refinement industries. The environmental application contains over 40 EPA listed VOCs and is compliant with EPA method 8240. Additionally, with our fully loaded gas chromatograph rental instruments and years of experience, we can configure a Voyager for almost any custom portable GC application.
Ranges
- All Gases
- The Voyager GC is capable of detecting a large number of gases. Please call for ranges on compounds not in this list.
- Acetone (assay 1)
- 0.015 to 100 ppm
- Acrylonitrile (assay 4)
- 3 to 180 ppm
- t-Amyl methyl ether TAME (assay 2)
- 0.1 to 100 ppm
- Benzene (assay 1 and assay 2)
- 0.005 to 35 ppm (assay 1)
- 0.005 to 100 ppm (assay 2)
- Bromodichlorometane (assay 1)
- 0.006 to 15 ppm
- Bromoform (assay 1)
- 0.01 to 25 ppm
- Bromomethane (assay 1)
- 0.025 to 100 ppm
- Butadiene (assay 4)
- 0.1 to 100 ppm
- Carbon Disulfide (assay 1)
- 0.35 to 200 ppm
- Carbon Tetrachloride (assay 1)
- 0.01 to 20 ppm
- Chlorobenzene (assay 1)
- 0.01 to 25 ppm
- Choroethane (assay 1)
- 0.5 to 200 ppm
- 2-Chloroethyl Vinyl Ether (assay 1)
- 0.05 to 150 ppm
- Chloroform (assay 1)
- 1 to 250 ppm
- Chloromethane (assay 1)
- 0.75 to 200 ppm
- Dibromochloromethane (assay 1)
- 0.075 to 200 ppm
- 1,2-Dichlorobenzene (assay 1)
- 0.03 to 50 ppm
- 1,3-Dichlorobenzene (assay 1)
- 0.03 to 50 ppm
- 1,4-Dichlorobenzene (assay 1)
- 0.03 to 50 ppm
- 1,2-Dichloroethane (assay 1)
- 1.5 to 250 ppm
- 1,1-Dichloroethylene (assay 1)
- 0.25 to 200 ppm
- c-1,2-Dichloroethylene (assay 1)
- 0.01 to 60 ppm
- t-1,2-Dichloroethylene (assay 1)
- 0.01 to 100 ppm
- 1,2-Dichloropropane (assay 1)
- 0.1 to 200 ppm
- c-1,3-Dichloropropene (assay 1)
- 0.025 to 20 ppm
- t-1,3-Dichloropropene (assay 1)
- 0.03 to 125 ppm
- Ethylbenzene (assay 1 and assay 2)
- 0.01 to 100 ppm (assay 1)
- 0.01 to 150 ppm (assay 2)
- 2-Hexanone (assay 1)
- 0.045 to 100 ppm
- Methanol (assay 2)
- 10 to 500 ppm
- Methyl t-butyl ether MTBE (assay 2)
- 0.1 to 100 ppm
- Methyl Ethyl Ketone (assay 1)
- 0.015 to 100 ppm
- Methyl Isobutyl Ketone (assay 1)
- 0.02 to 100 ppm
- Methylene Chloride (assay 1)
- 0.05 to 200 ppm
- Sytrene (assay 1 and assay 4)
- 0.045 to 60 ppm (assay 1)
- 0.1 to 500 ppm (assay 4)
- 1,1,2,2-Tetrachloroethane (assay 1)
- 0.3 to 200 ppm
- Tetrachloroethylene (assay 1)
- 0.01 to 35 ppm
- Toluene (assay 1 and assay 2)
- 0.01 to 100 ppm (assay 1)
- 0.01 to 200 ppm (assay 2)
- 1,1,1-Trichloroethane (assay 1)
- 1.5 to 250 ppm
- 1,1,2-Trichloroethane (assay 1)
- 0.1 to 200 ppm
- Trichloroethylene (assay 1)
- 0.007 to 250 ppm
- Vinyl Acetate (assay 1)
- 0.07 to 150 ppm
- Vinyl Chloride (assay 1)
- 0.025 to 150 ppm
- m-Xylene (assay 1 and assay 2)
- 0.01 to 150 ppm
- o-Xylene (assay 1 and assay 2)
- 0.012 to 150 ppm
- p-Xylene (assay 1)
- 0.01 to 150 ppm
Features
- Fully self contained portable GC
- Assay 1 comes preconfigured for 40 EPA listed VOC’s
- Compliant with EPA methods 8240 and TO-14
- Assay 2 for petrochemical refining
- Assay 4 for ABS rubber
- Customs configurations available
- Large graphical backlit LCD display
- View chromatograms or tabular results in the field
- Optional dual PID/ECD detector
- 3 column design with isothermal oven
- PC interface
- Built in sampling pump
- Onboard carrier gas cylinder
- Rechargeable battery
- Datalogging
- PC interface
Specifications
- Datalogging capacity 48 sets of results
- PC interface RS232
- Detector PID and optional ECD
- Input power rechargeable battery, 10 to 18 VDC, or 115/230 VAC
- Runtime 8 hours
- Operating temperature 41 to 105°F
- Operating humidity 0 to 100% RH
- Dimensions 15.4″ x 10.6″ x 5.9″
- Weight 15 lbs